31 Appendix 5, Regulation (EU) 2017/1151

In forcesince 1 Sep 2023

This appendix sets out the specifications and calibration of PEMS components and signals

A | — | undiluted CO2 concentration [%]

a0 | — | y-axis intercept of the linear regression line

a1 | — | slope of the linear regression line

B | — | diluted CO2 concentration [%]

C | — | diluted NO concentration [ppm]

c | — | analyser response in the oxygen interference test

Cb | | Measured diluted NO concentration through bubbler

cFS,b | — | full scale HC concentration in step (b) [ppmC1]

cFS,d | — | full scale HC concentration in step (d) [ppmC1]

cHC(w/NMC) | — | HC concentration with CH4 or C2H6 flowing through the NMC [ppmC1]

cHC(w/o NMC) | — | HC concentration with CH4 or C2H6 bypassing the NMC [ppmC1]

cm,b | — | measured HC concentration in step (b) [ppmC1]

cm,d | — | measured HC concentration in step (d) [ppmC1]

cref,b | — | reference HC concentration in step (b) [ppmC1]

cref,d | — | reference HC concentration in step (d) [ppmC1]

D | — | undiluted NO concentration [ppm]

De | — | expected diluted NO concentration [ppm]

E | — | absolute operating pressure [kPa]

ECO2 | — | per cent CO2 quench

E(dp) | — | PEMS-PN analyser efficiency

EE | — | ethane efficiency

EH2O | — | per cent water quench

EM | — | methane efficiency

EO2 | — | oxygen interference

F | — | water temperature [K]

G | — | saturation vapour pressure [kPa]

H | — | water vapour concentration [%]

Hm | — | maximum water vapour concentration [%]

NOX,dry | — | moisture-corrected mean concentration of the stabilized NOX recordings

NOX,m | — | mean concentration of the stabilized NOX recordings

NOX,ref | — | reference mean concentration of the stabilized NOX recordings

r2 | — | coefficient of determination

t0 | — | time point of gas flow switching [s]

t10 | — | time point of 10 % response of the final reading

t50 | — | time point of 50 % response of the final reading

t90 | — | time point of 90 % response of the final reading

Tbd | — | to be determined

X | — | independent variable or reference value

xmin | — | minimum value

Y | — | dependent variable or measured value

The accuracy and linearity of analysers, flow-measuring instruments, sensors and signals shall be traceable to international or national standards. Any sensors or signals that are not directly traceable (e.g., simplified flow-measuring instruments) shall be calibrated alternatively against chassis dynamometer laboratory equipment that has been calibrated against international or national standards.

All analysers, flow-measuring instruments, sensors and signals shall comply with the linearity requirements given in Table A5/1. If air flow, fuel flow, the air-to-fuel ratio or the exhaust mass flow rate is obtained from the ECU, the calculated exhaust mass flow rate shall meet the linearity requirements specified in Table A5/1.

[Optional to determine exhaust mass flow.] [Optional parameter.] [The linearity check shall be verified with soot-like particles, as these are defined in paragraph 6.2. of this appendix.] [To be updated based on error propagation and traceability charts.]

Measurement parameter/instrument | xmin a1 1 a0 |

Slope

a1

Standard error of the estimate

SEE

Coefficient of determination r2

Fuel flow rate | ≤ 1 % xmax | 0,98 – 1,02 | ≤ 2 % of xmax | ≥ 0,990

Air flow rate (15) | ≤ 1 % xmax | 0,98 – 1,02 | ≤ 2 % of xmax | ≥ 0,990

Exhaust mass flow rate | ≤ 2 % xmax | 0,97 – 1,03 | ≤ 3 % of xmax | ≥ 0,990

Gas analysers | ≤ 0,5 % max | 0,99 – 1,01 | ≤ 1 % of xmax | ≥ 0,998

Torque | ≤ 1 % xmax | 0,98 – 1,02 | ≤ 2 % of xmax | ≥ 0,990

PN analysers | ≤ 5 % xmax | 0,85 – 1,15 | ≤ 10 % of xmax | ≥ 0,950

The linearity requirements pursuant to paragraph 3.2. shall be verified:

(a)

for each gas analyser at least every 12 months or whenever a system repair or component change or modification is made that could influence the calibration;

(b)

for other relevant instruments, such as PN analysers, exhaust mass flow meters and traceably calibrated sensors, whenever damage is observed, as required by internal audit procedures or by the instrument manufacturer but no longer than one year before the actual test.

The linearity requirements pursuant to paragraph 3.2. for sensors or ECU signals that are not directly traceable shall be performed using a measurement device with a traceable calibration on the chassis dynamometer, once for each PEMS-vehicle setup.

The relevant analysers, instruments and sensors shall be brought to their normal operating condition according to the recommendations of their manufacturer. The analysers, instruments and sensors shall be operated at their specified temperatures, pressures and flows.

The linearity shall be verified for each normal operating range by executing the following steps:

(a)

The analyser, flow-measuring instrument or sensor shall be set to zero by introducing a zero signal. For gas analysers, purified synthetic air or nitrogen shall be introduced to the analyser port via a gas path that is as direct and short as possible.

(b)

The analyser, flow-measuring instrument or sensor shall be spanned by introducing a span signal. For gas analysers, an appropriate span gas shall be introduced to the analyser port via a gas path that is as direct and short as possible.

(c)

The zero procedure of (a) shall be repeated.

(d)

The linearity shall be verified by introducing at least 10, approximately equally spaced and valid, reference values (including zero). The reference values with respect to the concentration of components, the exhaust mass flow rate or any other relevant parameter shall be chosen to match the range of values expected during the emissions test. For measurements of exhaust mass flow, reference points below 5 per cent of the maximum calibration value can be excluded from the linearity verification.

(e)

For gas analysers, known gas concentrations in accordance with paragraph 5. shall be introduced to the analyser port. Sufficient time for signal stabilisation shall be given. For particle number analysers, the particle number concentrations shall be at least two times the limit of detection (defined in point 6.2).

(f)

The values under evaluation and, if needed, the reference values shall be recorded at a constant frequency which is a multiple of 1.0 Hz over a period of 30 seconds (60 s for particle number analysers).

(g)

The arithmetic mean values over the 30 (or 60 s) seconds period shall be used to calculate the least squares linear regression parameters, with the best-fit equation having the form:

ya1x a0

where:

y is the actual value of the measurement system

a1 is the slope of the regression line

x is the reference value

a0 is the y intercept of the regression line

The standard error of estimate (SEE) of y on x and the coefficient of determination (r2) shall be calculated for each measurement parameter and system.

(h)

The linear regression parameters shall meet the requirements specified in Table A5/1.

Non-traceable flow-measuring instruments, sensors or ECU signals, that cannot directly be calibrated according to traceable standards, shall be calibrated on a chassis dynamometer. The procedure shall follow, as far as applicable, the requirements of UN Regulation No 154. If necessary, the instrument or sensor to be calibrated shall be installed on the test vehicle and operated according to the requirements of Appendix 4. The calibration procedure shall follow whenever possible the requirements of paragraph 3.4.2. At least 10 appropriate reference values shall be selected as to ensure that at least 90 per cent of the maximum value expected to occur during the RDE test is covered.

If a non-traceable flow-measuring instrument, sensor or ECU signal for determining exhaust flow is to be calibrated, a reference exhaust mass flow meter with traceable calibration or the CVS shall be attached to the vehicle’s tailpipe. It shall be ensured that the vehicle exhaust is accurately measured by the exhaust mass flow meter according to paragraph 3.4.3. of Appendix 4. The vehicle shall be operated by applying constant throttle at a constant gear selection and chassis dynamometer load.

The gaseous components shall be measured with analysers specified in paragraph 4.1.4., Annex B5 to UN Regulation No 154. If an NDUV analyser measures both NO and NO2, a NO2/NO converter is not required.

Any analyser not meeting the design specifications of paragraph 4.1.1. is permissible provided that it fulfils the requirements of paragraph 4.2. The manufacturer shall ensure that the alternative analyser achieves an equivalent or higher measurement performance compared to a standard analyser over the range of pollutant concentrations and co-existing gases that can be expected from vehicles operated with permissible fuels under moderate and extended conditions of valid RDE testing as specified in paragraphs 5., 6. and 7. of this Appendix. Upon request, the manufacturer of the analyser shall submit in writing supplemental information, demonstrating that the measurement performance of the alternative analyser is consistently and reliably in line with the measurement performance of standard analysers. Supplemental information shall contain:

(a)

a description of the theoretical basis and the technical components of the alternative analyser;

(b)

a demonstration of equivalency with the respective standard analyser specified in paragraph 4.1.1. over the expected range of pollutant concentrations and ambient conditions of the type-approval test defined in UN Regulation No 154 as well as a validation test as described in paragraph 3. of Appendix 6 for a vehicle equipped with a spark-ignition and compression-ignition engine; the manufacturer of the analyser shall demonstrate the significance of equivalency within the permissible tolerances given in paragraph 3.3. of Appendix 6.

(c)

a demonstration of equivalency with the respective standard analyser specified in paragraph 4.1.1. with respect to the influence of atmospheric pressure on the measurement performance of the analyser; the demonstration test shall determine the response to span gas having a concentration within the analyser range to check the influence of atmospheric pressure under moderate and extended altitude conditions defined in paragraph 5.2. Such a test can be performed in an altitude environmental test chamber.

(d)

a demonstration of equivalency with the respective standard analyser specified in paragraph 4.1.1. over at least three on-road tests that fulfil the requirements of this Appendix.

(e)

a demonstration that the influence of vibrations, accelerations and ambient temperature on the analyser reading does not exceed the noise requirements for analysers set out in paragraph 4.2.4.

Approval authorities may request additional information to substantiate equivalency or refuse approval if measurements demonstrate that an alternative analyser is not equivalent to a standard analyser.

In addition to the linearity requirements defined for each analyser in paragraph 3., the compliance of analyser types with the specifications laid down in paragraphs 4.2.2. to 4.2.8. shall be demonstrated by the analyser manufacturer. Analysers shall have a measuring range and response time appropriate to measure with adequate accuracy the concentrations of the exhaust gas components at the applicable emissions standard under transient and steady state conditions. The sensitivity of the analysers to shocks, vibration, aging, variability in temperature and air pressure as well as electromagnetic interferences and other impacts related to vehicle and analyser operation shall be limited as far as possible.

The accuracy, defined as the deviation of the analyser reading from the reference value, shall not exceed 2 per cent of reading or 0.3 per cent of full scale, whichever is larger.

The precision, defined as 2.5 times the standard deviation of 10 repetitive responses to a given calibration or span gas, shall be no greater than 1 per cent of the full scale concentration for a measurement range equal or above 155 ppm (or ppmC1) and 2 per cent of the full scale concentration for a measurement range of below 155 ppm (or ppmC1).

The noise shall not exceed 2 per cent of full scale. Each of the 10 measurement periods shall be interspersed with an interval of 30 seconds in which the analyser is exposed to an appropriate span gas. Before each sampling period and before each span period, sufficient time shall be given to purge the analyser and the sampling lines.

The drift of the zero response, defined as the mean response to a zero gas during a time interval of at least 30 seconds, shall comply with the specifications given in Table A5/2.

The drift of the span response, defined as the mean response to a span gas during a time interval of at least 30 seconds, shall comply with the specifications given in Table A5/2.

Pollutant | Absolute Zero response drift | Absolute Span response drift

CO2 | ≤ 1000 ppm over 4 h | ≤ 2 % of reading or ≤ 1000 ppm over 4 h, whichever is larger

CO | ≤ 50 ppm over 4 h | ≤ 2 % of reading or ≤ 50 ppm over 4 h, whichever is larger

PN | 5000 particles per cubic centimetre over 4 h | According to manufacturer specifications

NOX | ≤ 3 ppm over 4 h | ≤ 2 % of reading or 3 ppm over 4 h, whichever is larger

CH4 | ≤ 10 ppm C1 | ≤ 2 % of reading or ≤ 10 ppm C1 over 4 h, whichever is larger

THC | ≤ 10 ppm C1 | ≤ 2 % of reading or ≤ 10 ppm C1 over 4 h, whichever is larger

The rise time, defined as the time between the 10 per cent and 90 per cent response of the final reading (t10 to t90; see paragraph 4.4.), shall not exceed 3 seconds.

Exhaust gases may be measured wet or dry. A gas-drying device, if used, shall have a minimal effect on the composition of the measured gases. Chemical dryers are not permitted.

The provisions in paragraphs 4.3.2. to 4.3.5. define additional performance requirements for specific analyser types and apply only to cases in which the analyser under consideration is used for RDE emission measurements.

If a NOX converter is applied, for example to convert NO2 into NO for analysis with a chemiluminescence analyser, its efficiency shall be tested by following the requirements in paragraph 5.5. of Annex B5 to UN Regulation No 154. The efficiency of the NOX converter shall be verified no longer than one month before the emissions test.

(a)

Optimization of the detector response

If hydrocarbons are measured, the FID shall be adjusted as specified by the instrument manufacturer. A propane-in-air or propane-in-nitrogen span gas shall be used to optimize the response in the most common operating range.

(b)

Hydrocarbon response factors

If hydrocarbons are measured, the hydrocarbon response factor of the FID shall be verified by following the provisions of paragraph 5.4.3. of Annex B5 to UN Regulation No 154, using propane-in-air or propane-in-nitrogen as span gases and purified synthetic air or nitrogen as zero gases, respectively.

(c)

Oxygen interference check

The oxygen interference check shall be performed when introducing a FID into service and after major maintenance intervals. A measuring range shall be chosen in which the oxygen interference check gases fall in the upper 50 per cent. The test shall be conducted with the oven temperature set as required. The specifications of the oxygen interference check gases are described in paragraph 5.3.

The following procedure applies:

(i)

The analyser shall be set at zero;

(ii)

The analyser shall be spanned with a 0 per cent oxygen blend for positive ignition engines and a 21 per cent oxygen blend for compression ignition engines;

(iii)

The zero response shall be rechecked. If it has changed by more than 0.5 per cent of full scale, steps (i) and (ii) shall be repeated;

(iv)

The 5 per cent and 10 per cent oxygen interference check gases shall be introduced;

(v)

The zero response shall be rechecked. If it has changed by more than ±1 per cent of full scale, the test shall be repeated;

(vi)

The oxygen interference EO2 [%] shall be calculated for each oxygen interference check gas in step (iv) as follows:

EO2

cref,d c

cref,d

100

where the analyser response is:

c

cref,d cFS,b

cm,b

cm,b

cFS,d

where:

cref,b | | is the reference HC concentration in step (ii) [ppmC1]

cref,d | | is the reference HC concentration in step (iv) [ppmC1]

cFS,b | | is the full scale HC concentration in step (ii) [ppmC1]

cFS,d | | is the full scale HC concentration in step (iv) [ppmC1]

cm,b | | is the measured HC concentration in step (ii) [ppmC1]

cm,d | | is the measured HC concentration in step (iv) [ppmC1]

(vii)

The oxygen interference EO2 shall be less than ±1.5 per cent for all required oxygen interference check gases.

(viii)

If the oxygen interference EO2 is higher than ±1.5 per cent, corrective action may be taken by incrementally adjusting the air flow (above and below the manufacturer's specifications), the fuel flow and the sample flow.

(ix)

The oxygen interference check shall be repeated for each new setting.

If hydrocarbons are analysed, a NMC can be used to remove non-methane hydrocarbons from the gas sample by oxidizing all hydrocarbons except methane. Ideally, the conversion for methane is 0 per cent and for the other hydrocarbons, represented by ethane, is 100 per cent. For the accurate measurement of NMHC, the two efficiencies shall be determined and used for the calculation of the NMHC emissions (see paragraph 6.2. of Appendix 7). It is not necessary to determine the methane conversion efficiency in the case where the NMC-FID is calibrated according to method (b) in paragraph 6.2. of Appendix 7 by passing the methane/air calibration gas through the NMC.

(a)

Methane conversion efficiency

Methane calibration gas shall be flowed through the FID with and without bypassing the NMC; the two concentrations shall be recorded. The methane efficiency shall be determined as:

EM1

CHCw/NMC

CHCw/o NMC

where:

cHC(w/NMC) | | is the HC concentration with CH4 flowing through the NMC [ppmC1]

cHC(w/o NMC) | | is the HC concentration with CH4 bypassing the NMC [ppmC1]

(b)

Ethane conversion efficiency

Ethane calibration gas shall be flowed through the FID with and without bypassing the NMC; the two concentrations shall be recorded. The ethane efficiency shall be determined as:

EE1

CHCw/NMC

CHCw/o NMC

where:

cHC(w/NMC) | | is the HC concentration with C2H6 flowing through the NMC [ppmC1]

cHC(w/o NMC) | | is the HC concentration with C2H6 bypassing the NMC [ppmC1]

(a)

General

Other gases than the ones being analysed can affect the analyser reading. A check for interference effects and the correct functionality of analysers shall be performed by the analyser manufacturer prior to market introduction at least once for each type of analyser or device addressed in paragraphs 4.3.5. (b) to (f).

(b)

CO analyser interference check

Water and CO2 can interfere with the measurements of the CO analyser. Therefore, a CO2 span gas, having a concentration of 80 to 100 per cent of the full scale of the maximum operating range of the CO2 analyser used during the test, shall be bubbled through water at room temperature and the analyser response recorded. The analyser response shall not be more than 2 per cent of the mean CO concentration expected during normal on-road testing or ± 50 ppm, whichever is larger. The interference check for H2O and CO2 may be run as separate procedures. If the H2O and CO2 levels used for the interference check are higher than the maximum levels expected during the test, each observed interference value shall be scaled down by multiplying the observed interference with the ratio of the maximum expected concentration value during the test and the actual concentration value used during this check. Separate interference checks with concentrations of H2O that are lower than the maximum concentration expected during the test may be run and the observed H2O interference shall be scaled up by multiplying the observed interference with the ratio of the maximum H2O concentration value expected during the test and the actual concentration value used during this check. The sum of the two scaled interference values shall meet the tolerance specified in this point.

(c)

NOX analyser quench check

The two gases of concern for CLD and HCLD analysers are CO2 and water vapour. The quench response to these gases is proportional to the gas concentrations. A test shall determine the quench at the highest concentrations expected during the test. If the CLD and HCLD analysers use quench compensation algorithms that utilize H2O or CO2 measurement analysers or both, quench shall be evaluated with these analysers active and with the compensation algorithms applied.

(i)

CO2 quench check

A CO2 span gas having a concentration of 80 to 100 per cent of the maximum operating range shall be passed through the NDIR analyser; the CO2 value shall be recorded as A. The CO2 span gas shall then be diluted by approximately 50 per cent with NO span gas and passed through the NDIR and CLD or HCLD; the CO2 and NO values shall be recorded as B and C, respectively. The CO2 gas flow shall then be shut off and only the NO span gas shall be passed through the CLD or HCLD; the NO value shall be recorded as D. The per cent quench shall be calculated as:

ECO21

C A

D A D B

100

where:

A | | is the undiluted CO2 concentration measured with the NDIR [%]

B | | is the diluted CO2 concentration measured with the NDIR [%]

C | | is the diluted NO concentration measured with the CLD or HCLD [ppm]

D | | is the undiluted NO concentration measured with the CLD or HCLD [ppm]

Alternative methods of diluting and quantifying of CO2 and NO span gas values such as dynamic mixing/blending are permitted upon approval of the approval authority.

(ii)

Water quench check

This article is long: the rest is at the official publication.

  1. 1 Sep 2023 – todayin force
  2. 1 Jan 2019 – 1 Sep 2023earlier wording
  3. 27 Jul 2017 – 1 Jan 2019earlier wording
  4. 1 Jun 2017 – 27 Jul 2017earlier wording