The following three parameters are critical to ensure obtaining reliable results with the luminometer:
having a sufficient sensitivity giving a stable background in control wells;
having no gradient over the plate due to long reading times; and
having no light contamination in adjacent wells from strongly active wells.
Prior to testing it is recommended to ensure having appropriate luminescence measurements, by testing a control plate set-up as described below (triplicate analysis).
1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12
A | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO
B | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO
C | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO
D | EGDMA 0,98 | EGDMA 1,95 | EGDMA 3,9 | EGDMA 7,8 | EGDMA 15,6 | EGDMA 31,25 | EGDMA 62,5 | EGDMA 125 | EGDMA 250 | EGDMA 500 | EGDMA 1000 | EGDMA 2000
E | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO
F | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO
G | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO
H | DMSO | DMSO | DMSO | DMSO | DMSO | DMSO | CA 4 | CA 8 | CA 16 | CA 32 | CA 64 | Blank
EGDMAEthylene glycol dimethacrylate (CAS No.: 97-90-5) a strongly inducing chemical
CACinnamic aldehyde, positive reference (CAS No.: 104-55-2)
The quality control analysis should demonstrate:
a clear dose-response in row D, with the Imax > 20 fold above background (in most cases Imax values between 100 and 300 are reached);
no dose-response in row C and E (no induction value above 1,5 (ideally not above 1,3) due to possible light contamination especially next to strongly active wells in the EGDMA row;
no statistically significant difference between the rows A, B, C, E, F and G. (i.e. no gradient over plate); and
variability in any of the rows A, B, C, E, F and G and in the DMSO wells in row H should be below 20 % (i.e. stable background).