Three sets of reference controls (i.e. samples constituted only by the peptide dissolved in the appropriate solvent) should be included in the analysis sequence:

Reference control A: used to verify the suitability of the HPLC system.

Reference control B: included at the beginning and at the end of the analysis sequence to verify stability of reference controls over the analysis time.

Reference control C: included in the analysis sequence to verify that the solvent used to dissolve the test chemical does not impact the percent peptide depletion.

Calibration standards and reference controls |

STD1

STD2

STD3

STD4

STD5

STD6

Dilution buffer

Reference control A, rep 1

Reference control A, rep 2

Reference control A, rep 3

Co-elution controls |

Co-elution control 1 for test chemical 1

Co-elution control 2 for test chemical 2

Reference controls |

Reference control B, rep 1

Reference control B, rep 2

Reference control B, rep 3

First set of replicates |

Reference control C, rep 1

Cinnamic aldehyde, rep 1

Sample 1, rep 1

Sample 2, rep 1

Second set of replicates |

Reference control C, rep 2

Cinnamic aldehyde, rep 2

Sample 1, rep 2

Sample 2, rep 2

Third set of replicates |

Reference control C, rep 3

Cinnamic aldehyde, rep 3

Sample 1, rep 3

Sample 2, rep 3

Reference controls |

Reference control B, rep 4

Reference control B, rep 5

Reference control B, rep 6