Three sets of reference controls (i.e. samples constituted only by the peptide dissolved in the appropriate solvent) should be included in the analysis sequence:
Reference control A: used to verify the suitability of the HPLC system.
Reference control B: included at the beginning and at the end of the analysis sequence to verify stability of reference controls over the analysis time.
Reference control C: included in the analysis sequence to verify that the solvent used to dissolve the test chemical does not impact the percent peptide depletion.
Calibration standards and reference controls |
STD1
STD2
STD3
STD4
STD5
STD6
Dilution buffer
Reference control A, rep 1
Reference control A, rep 2
Reference control A, rep 3
Co-elution controls |
Co-elution control 1 for test chemical 1
Co-elution control 2 for test chemical 2
Reference controls |
Reference control B, rep 1
Reference control B, rep 2
Reference control B, rep 3
First set of replicates |
Reference control C, rep 1
Cinnamic aldehyde, rep 1
Sample 1, rep 1
Sample 2, rep 1
Second set of replicates |
Reference control C, rep 2
Cinnamic aldehyde, rep 2
Sample 1, rep 2
Sample 2, rep 2
Third set of replicates |
Reference control C, rep 3
Cinnamic aldehyde, rep 3
Sample 1, rep 3
Sample 2, rep 3
Reference controls |
Reference control B, rep 4
Reference control B, rep 5
Reference control B, rep 6