The relevant requirements of Annex I, the specific requirements of this Annex and the conformity assessment procedures listed in this Annex apply to measuring systems intended for the continuous and dynamic measurement of quantities (mass and, if applicable, energy) of compressed gases, (CG).
Such measuring systems shall not be considered as utility measuring instruments as referred to in Annex I.
Meter | An instrument designed to measure continuously and to ensure the memorisation and display of the quantity of gas, at metering conditions, flowing through the measurement transducer in a closed, fully charged conduit.
Calculator | A part of a meter that receives the output signals from the measurement transducers, and possibly from associated measuring instruments, and displays the measurement results.
Associated measuring instrument | An instrument connected to the calculator for measuring certain quantities, which are characteristic of the gas, with a view to making a correction and/or conversion.
Conversion device | A part of the calculator, which by taking into account the characteristics of the gas, automatically converts the mass of the gas into the amount of energy delivered or received.
Measuring system | A system that comprises, in addition to the meter itself, a transfer point, gas piping and all devices required to ensure correct measurement or intended to facilitate the measuring operations.
Compressed gas dispenser | A system intended for the fuelling of vehicles (such as motor vehicles, rail engines, boats, vessels and aircraft) with compressed gaseous fuel.
Transfer point | Physical location at which the gas is defined as being delivered or received.
Self-service arrangement | An arrangement that allows customers to use a measuring system for the purpose of obtaining gas for their own use.
Self-service device | A specific device that is part of a self-service arrangement and that allows one or more measuring systems to perform in that self-service arrangement.
Minimum measured quantity (“MMQ”) | The smallest quantity of gas for which the measurement is metrologically acceptable for the measuring system.
Direct indication |
The indication of mass and, if applicable, energy, corresponding to the measure that the meter is physically capable of measuring.
Note:
The direct indication may be converted into another quantity using a conversion device.
Interruptible | A measuring system is considered as interruptible when the gas flow can be stopped easily and rapidly.
Non-interruptible | A measuring system is considered as non-interruptible when the gas flow cannot be stopped easily and rapidly.
Flowrate range | The range between the minimum flowrate (Qmin) and maximum flowrate (Qmax).
The manufacturer shall specify the rated operating conditions for the measuring system, in particular:
1.1. The flowrate range
The flowrate range is subject to the following conditions:
the flowrate range of a measuring system shall be within the flowrate range of each of its elements, in particular the meter; and
the ratio between the maximum and minimum flow rate shall be no less than 10.
1.2. The properties of the gas to be measured by the instrument by specifying the name, the type or the following relevant characteristics of that gas such as:
temperature range;
pressure range;
the calorific value of the gas;
the nature and characteristics of the gas to be measured.
1.3. The nominal value of the AC voltage supply and/or limits of the DC voltage supply.
2.1. The MPE on the indication of measured or converted amounts transferred at the transfer point is set out in Table 1.
Type of compressed gas measuring systems | Accuracy Class (MPE [% of measured value])
Compressed hydrogen measuring systems | 2,0
Other compressed gas measuring systems | 1,5
The MPE on the MMQ equals twice the value given in Table 1.
2.2. The MMQ of a measuring system shall have the form 1 × 10n, 2 × 10n, or 5 × 10n authorised units of mass or energy, where n is a positive or negative whole number, or zero.
The MMQ shall satisfy the conditions of use of the measuring system; except in exceptional cases, the measuring system shall not be used for measuring quantities less than that MMQ.
2.3. The measuring system shall not exploit the MPEs or systematically favour any party.
3.1. The effect of an electromagnetic disturbance on a measuring system shall be one of the following:
the change in the measurement result is not greater than the critical change value pursuant to point 3.2;
the indication of the measurement result shows a momentary variation that cannot be interpreted, memorised or transmitted as a measurement result; furthermore, in the case of an interruptible system, that can also mean the impossibility to perform any measurement; or
the change in the measurement result is greater than the critical change value pursuant to point 3.2, in which case the measuring system shall permit the retrieval of the measurement result just before the critical change value occurred and cut off the flow.
3.2. The critical change value is the greater of the following values:
10 % of the MPE,
3 % of the MMQ; in the case of a failure of the main power source, the critical change value shall be increased by 5 % of the MMQ.
For systems fitted with meters with moving parts, after an appropriate durability test, taking into account the period of time estimated by the manufacturer, has been performed, the following criterion shall be satisfied:
the variation of the measurement result after such durability test, when compared with the initial measurement result, shall not exceed two fifths of the MPE.
5.1. For any measured quantity relating to the same measurement, the indications and, if applicable, printouts provided by various devices, including those devices that are part of a self-service arrangement, shall have the same scale interval and the results shall not deviate from one another.
The scale interval of a CG measuring system shall not exceed 1,5 % of the MMQ.
5.2. It shall not be possible to divert the measured quantity in normal conditions of use unless it is readily apparent.
5.3. During the warm-up time of the CG measuring system, no measurements shall take place.
5.4.1. A measuring system for direct sales shall be provided with means for resetting the display to zero.
It shall not be possible to divert measured gas downstream of the meter during a filling operation.
5.4.2. The display of the quantity on which the transaction is based shall remain until all parties in the transaction have accepted the measurement result.
5.4.3. Measuring systems for direct sales shall be interruptible.
5.4.4. Measuring systems for direct sales shall display measurement results in units of mass and, if applicable, energy.
5.5.1. It shall not be possible to reset displays on the measuring systems to zero during a measurement.
5.5.2. The start of a new measurement shall be inhibited until the display has been reset to zero.
5.5.3. Where a measuring system is fitted with a price display, the difference between the indicated price and the price calculated from the unit price and the indicated quantity shall not exceed the smallest currency unit. However, that difference is not required to be less than the smallest monetary value.
A measuring system shall either be provided with an emergency power supply device that will safeguard all measuring functions during the failure of the main power supply device or be equipped with means to save and display the data present in order to permit the conclusion of the transaction in progress and with means to stop the flow of gas at the moment of failure of the main power supply device.
The metered quantity shall be displayed in kilograms, or their decimal multiples or submultiples, and, if applicable, in joules or watt-hours, or their decimal multiples.
The conformity assessment procedures referred to in Article 17 that the manufacturer can choose between are:
B + F or B + D or H1 or G.’.