As regards the SAR measurements to demonstrate equipment compliance, harmonised standard EN 50566:2017 provides that they are to be performed according to EN 62209-2:2010, Clauses 5 and 6. EN 62209-2:2010 does not, however, set out a measurement methodology for mobile communication devices featuring certain advanced functionalities that can be relevant as regards limb SAR measurement. Since the iPhone 12 A2403 has this kind of features, the development of a specific testing protocol was necessary.
80 4. COMMISSION’S ASSESSMENT › 4.3. Substantive points › 4.3.2. SAR limits, Decision_act (EU) 2025/1760
In forcesince 19 Aug 2025