According to available scientific evidence on equipment intended for brain stimulation that apply electrical currents or magnetic or electromagnetic fields that penetrate the cranium to modify neuronal activity in the brain as referred to in Section 6 of Annex XVI to Regulation (EU) 2017/745, such as those for transcranial magnetic stimulation or transcranial electric stimulation, the use of such products may cause side effects, for example, atypical brain development, abnormal patterns of brain activity, increase metabolic consumption, fatigue, anxiety, irritability, headaches, muscle twitches, tics, seizures, vertigo and skin irritation at the electrode site. While such equipment is not surgically invasive, the electrical currents or magnetic or electromagnetic fields do penetrate the cranium to modify neuronal activity in the brain. Such modifications can have long-lasting effects and any unintended effects may be difficult to reverse. Such products should therefore be classified as Class III.
Recital 7, Regulation (EU) 2022/2347
In force