According to available scientific evidence on equipment intended to be used to reduce, remove or destroy adipose tissue, as referred to in Section 4 of Annex XVI to Regulation (EU) 2017/745, such as equipment for liposuction, radiofrequency lipolysis, ultrasound lipolysis, cryolipolyisis, laser lipolysis, infrared and electrical stimulation lipolysis, acoustic shockwave therapy or lipoplasty, the use of such products may cause side effects, for example, local inflammation, erythema, bruising and swelling. Side effects are often indicated as transient, but important and long-lasting effects are also reported, such as paradoxical adipose hyperplasia after cryolipolysis treatment. Those products should therefore be classified as class IIb. Such classification also corresponds to classification of active therapeutic devices whose functioning and risks profile are similar to those of the equipment without an intended medical purpose in question, intended to administer energy to or exchange energy with the human body or supply energy that will be absorbed by the human body in a potentially hazardous way, taking account of the nature, the density and site of application of the energy.
Recital 6, Regulation (EU) 2022/2347
In force