Equipment used in production or control operations shall be suitable for its intended purpose and shall not present any hazard to the product. The parts of production equipment that come into contact with the product shall not have any unwanted reactive, additive, adsorptive or absorptive properties that may affect the quality of the product.
Equipment that is critical to the quality of the products shall be subject to appropriate qualification.
Balances and measuring equipment shall be of an appropriate range and precision to ensure the accuracy of weighing operations.
Equipment shall be operated and maintained in such a way as to minimise the risk of error and to avoid contamination, cross-contamination and, in general, any adverse effect on the quality of the product.
Equipment shall be calibrated, inspected or checked, as appropriate, at defined intervals to ensure an adequate performance. In the case of computerised systems, the checks shall include an evaluation of the ability of the system to ensure data integrity. Appropriate records of those checks shall be maintained. Additional requirements relevant to the use of computerised systems are laid down in Annex IV.
Equipment shall be adequately cleaned to avoid the risk of contamination for the products. The cleaning or decontamination procedures shall be detailed in writing, ensuring that the cleaning equipment does not become a source of contamination. Equipment shall only be stored in a clean and dry condition.
The location and installation of the equipment shall be adequate to minimise risks of errors or contamination. In general, equipment, including laboratory equipment, shall not be moved between high-risk areas. If equipment is moved between high-risk areas, appropriate measures shall be applied to avoid the risk of cross-contamination. Where appropriate, the qualification status of the equipment moved shall also be reconsidered.
Fixed pipework shall be clearly labelled to indicate the content and, where applicable, the direction of flow.
Water for pharmaceutical use and, where appropriate, other water pipes shall be sanitised according to written procedures that detail the action limits for microbiological contamination and the measures to be taken.
Defective equipment shall be removed from production and quality control areas or, if the removal thereof is not possible, it shall be clearly labelled as defective.