This Annex applies to devices intended for detection or quantification of markers of Epstein-Barr virus (EBV) infection.
Table 1 applies to first-line assays for IgG antibodies against viral capsid antigen of EBV (anti-EBV VCA IgG).
Table 2 applies to qualitative and quantitative NAT devices for EBV DNA.
[Including testing of other EBV markers and parameters (e.g. VCA-IgM, EBNA-1 IgG, immunoblot) or previous / follow-up specimens to assess the true specimen status.] [Supplementary testing to confirm recent EBV infection: e.g. VCA-IgM, IgG-avidity, immunoblot analysis.] [At an assumed EBV prevalence of 80 % corresponding to an initial number of 1000 donors.] [Including pre-transplant recipients.]
Performance characteristic | Specimens | Specimen number, features, use | Acceptance criteria
Diagnostic sensitivity | Positive specimens |
≥400
including specimens from recent and past EBV infection,
low and high positive titre specimens
≥ 99% for confirmable past infection; overall sensitivity including recent infection shall be at least equivalent to the comparator device
Seroconversion panels | To be tested when available | diagnostic sensitivity during seroconversion shall represent the state of the art
Analytical sensitivity | Standards | International reference reagents, when available |
Diagnostic specificity | Negative specimens | ≥ 200 EBV negatives from unselected donors as compared to another EBV device | ≥ 99%
Hospitalised patients | ≥200 | Potential limitations for specificity, if any, shall be identified
Cross-reactivity | Potentially cross-reacting specimens |
≥100 in total
(e.g. RF+, related viruses or other infectious agents, pregnant women, etc.)
For target sequence amplification devices, a functionality control for each specimen (internal control) shall reflect the state of the art. This control shall as far as possible be used throughout the whole process, i.e. extraction, amplification/hybridisation, detection.
Genotype and/or subtype detection shall be demonstrated by appropriate primer or probe design validation and shall also be validated by testing characterised genotyped specimens.
Potential cross-reactivity of non-target nucleic acid sequences shall be analysed by appropriate primer or probe design validation and shall also be validated by testing selected specimens.
Results of quantitative NAT devices shall be traceable to international standards or calibrated reference materials, if available, and be expressed in international units utilised in the specific field of application.
[Reference: European Pharmacopeia 9.0, 2.6.21 Nucleic acid amplification techniques, Validation.]
Performance characteristic | Specimens | Specimen numbers, features, use | Acceptance criteria
Analytical sensitivity | WHO First International Standard Human EBV DNA (09/260; 5000000 IU/vial) (or calibrated reference materials) |
NAT sensitivity and NAT LOD shall be validated by dilution series of reference materials, testing of replicates (minimum 24) at different analyte concentrations, including those with transition from positive to negative results with the respective NAT device.
LOD shall be expressed as 95% positive cut-off value (IU/ml) after statistical analysis (e.g. Probit).
quantitative NAT: definition of lower, upper quantification limit, precision, accuracy,
‘linear’ measuring range, ‘dynamic range’.
Reproducibility at different concentration levels
According to the state of the art
Diagnostic sensitivity
EBV strain sensitivity
Patient specimens determined as EBV DNA positive by comparator device
Dilution series of EBV positive cell cultures may serve as potential substitutes
Qualitative NAT: ≥100
Quantitative NAT: ≥100
dilution series for demonstration of quantification efficiencies
Diagnostic specificity | Negative specimens |
Qualitative NAT: ≥500
Quantitative NAT: ≥100
According to the state of the art
Cross-reactivity | Potentially cross-reacting specimens |
≥20 specimens in total
Including human specimens positive for related human herpesviruses, e.g. CMV, HHV6, VZV
Herpesvirus positive cell cultures may serve as potential substitutes
According to the state of the art
Carry-over |
High EBV DNA positive;
EBV DNA negative
At least five runs with alternating high-positive and negative specimens shall be performed during robustness studies. The virus titres of the high positive specimens shall be representative of high virus titres occurring naturally. | According to the state of the art
Whole system failure rate | EBV DNA low-positive | ≥100 EBV DNA low-positive specimens shall be tested. These specimens shall contain a virus concentration equivalent to three times the 95 % positive cut-off virus concentration. | ≥99% positive