To ensure the safety and performance of the reprocessed single-use device, CS concerning risk management should include the procedure by which the reprocessing cycle is established. In particular, the reprocessing cycle should be based on the characteristics of the single-use device and the results of a technical assessment. To ensure that the performance and safety of the reprocessed single-use device remains equivalent to the original single-use device, it is necessary to determine a maximum number of reprocessing cycles which can be applied to the reprocessed single-use device while the performance and safety remain equivalent to the original single-use device.
Recital 4, Regulation (EU) 2020/1207
In forcesince 19 Aug 2020