Procedures for chemical disinfection referred to in point (e) of Article 11 shall cover the following, if applicable:
identification and concentration of chemicals required for the disinfection process;
contact time of the disinfectant;
temperature(s) to be used;
limits on temperature, concentration of solution(s), exposure time;
description of the accessories required for the disinfection process;
identification of water quality required;
techniques to be used including rinsing volume and time;
limits and monitoring of chemical residues remaining on the single-use device after disinfection;
limits and monitoring of chemical residues remaining on the single-use device from cleaning agents to ensure these residues do not interact adversely with the disinfectant;
criteria and/or requirements for approval or rejection.