Art. 51 Regulation (EU) 2017/745

Classification of devices

In forcesince 5 Apr 2017cited 10×
1.

Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. Classification shall be carried out in accordance with Annex VIII.

2.

Any dispute between the manufacturer and the notified body concerned, arising from the application of Annex VIII, shall be referred for a decision to the competent authority of the Member State in which the manufacturer has its registered place of business. In cases where the manufacturer has no registered place of business in the Union and has not yet designated an authorised representative, the matter shall be referred to the competent authority of the Member State in which the authorised representative referred to in the last indent of point (b) of the second paragraph of Section 2.2 of Annex IX has its registered place of business. Where the notified body concerned is established in a Member State other than that of the manufacturer, the competent authority shall adopt its decision after consultation with the competent authority of the Member State that designated the notified body.

The competent authority of the Member State in which the manufacturer has its registered place of business shall notify the MDCG and the Commission of its decision. The decision shall be made available upon request.

3.

At the request of a Member State the Commission shall after consulting the MDCG, decide, by means of implementing acts, on the following:

(a)

application of Annex VIII to a given device, or category or group of devices, with a view to determining the classification of such devices;

(b)

that a device, or category or group of devices, shall for reasons of public health based on new scientific evidence, or based on any information which becomes available in the course of the vigilance and market surveillance activities be reclassified, by way of derogation from Annex VIII.

4.

The Commission may also, on its own initiative and after consulting the MDCG, decide, by means of implementing acts, on the issues referred to in points (a) and (b) of paragraph 3.

5.

In order to ensure the uniform application of Annex VIII, and taking account of the relevant scientific opinions of the relevant scientific committees, the Commission may adopt implementing acts to the extent necessary to resolve issues of divergent interpretation and of practical application.

6.

The implementing acts referred to in paragraphs 3, 4 and 5 of this Article shall be adopted in accordance with the examination procedure referred to in Article 114(3).

  • Code de la sécurité socialer165-32Law

    … 4° Le nom du fabricant ; 5° Le type de produit ; 6° Le cas échéant, la classe du dispositif médical au sens de l'article 51 du règlement (UE) 2017/745 du 5 avril 2017. II.-La déclaration, prévue au même article L. 165-5, des …

  • Code de la santé publiquer5211-5Law

    … de sécurité du médicament et des produits de santé est l'autorité compétente mentionnée au paragraphe 2 de l'article 51 du règlement (UE) 2017/745 du 5 avril 2017 en matière de litige de classification.

  • Opinion of Advocate General Spielmann delivered on 11 September 2025. Dürr Dental SE v Cattani Deutschland Helmes GmbH & Co. KG.Decision

    Furthermore, devices are to be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks. [See Article 51(1) of Regulation 2017/745.] The conformity assessment procedure for …

    11 Sep 2025 · 62024CC0010

  • MPDG6Law

    … Verordnung (EU) 2017/745 oder der Verordnung (EU) 2017/746 sind der zuständigen Bundesoberbehörde nach Maßgabe von Artikel 51 Absatz 2 der Verordnung (EU) 2017/745 und Artikel 47 Absatz 2 der Verordnung (EU) 2017/746 zur Entscheidung …

  • Regulation (EU) 2021/22827Law

    … в) медицински изделия, класифицирани като клас IIb или III съгласно член 51 от Регламент (ЕС) 2017/745, за които съответните експертни групи …

  • BGH, EuGH-Vorlage vom 21. Dezember 2023 - I ZR 17/23Decision

    … Nach Art. 51 Abs. 1 der Verordnung (EU) 2017/745 werden Produkte unter Berücksichtigung ihrer Zweckbestimmung und der damit verbundenen Risiken in die Klassen I, IIa, IIb und III eingestuft und erfolgt die Klassifizierung gemäß Anhang …

    21 Dec 2023 · KORE305492023

  • Judgment of the Court (First Chamber) of 4 June 2026. Dürr Dental SE v Cattani Deutschland Helmes GmbH & Co. KG.Decision

    Article 51 of that regulation, headed ‘Classification of devices’, provides, in paragraphs 1 and 2 thereof: ‘1. Devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and …

    4 Jun 2026 · 62024CJ0010

  • Regulation (EU) 2022/2347Law

    … 1.], και ιδίως το άρθρο 51 παράγραφος 3 στοιχείο β), Εκτιμώντας τα ακόλουθα:

  • BGH, Beschluss vom 21. Dezember 2023 - I ZR 17/23Decision

    21 Dec 2023 · I_ZR__17-23

  • MPDG6Law