, The Genetically Modified Organisms (Contained Use) Regulations 2014

In forcesince 1 Oct 2014

Table 1a

Containment measures applicable to contained use involving micro-organisms in laboratories

1 “isolation” means, in relation to a laboratory, separation of the laboratory from other areas in the same building, or being in a separate building.

2 Entry must be through an airlock which is a chamber isolated from the laboratory. The clean side of the airlock must be separated from the restricted side by changing or showering facilities and preferably by interlocking doors.

3 Where viruses are not retained by the HEPA filters, extra requirements will be necessary for extract air.

4 Where the autoclave is outside the laboratory in which the contained use is being undertaken, but within the laboratory suite, there must be validated procedures for the safe transfer of material into that autoclave, which provide a level of protection equivalent to that which would be achieved by having an autoclave in that laboratory.

Containment Measures | Containment Levels

1 | 2 | 3 | 4

Facilities

1 | Laboratory suite: isolation1 | not required | not required | required | required

2 | Laboratory: sealable for fumigation | not required | not required | required | required

Equipment

3 | Surfaces impervious to water, resistant to acids, alkalis, solvents, disinfectants and decontamination agents and easy to clean | required for any bench | required for any bench | required for any bench and floor | required for any bench, floor, ceiling and walls

4 | Entry to laboratory via airlock2 | not required | not required | required where and to extent the risk assessment shows it is required | required

5 | Negative pressure relative to the pressure of the immediate surroundings | not required | not required | required except for activities where transmission does not occur by the airborne route | required

6 | Extract and input air from the laboratory must be HEPA filtered | not required | not required | HEPA filters required for extract air except for activities where transmission does not occur by the airborne route | HEPA filters required for input and extract air3

7 | Microbiological safety cabinet/enclosure | not required | required where and to extent the risk assessment shows it is required |

required, and all procedures with infective materials required to be contained within a cabinet/

enclosure

required,

and all procedures with infective materials required to be contained within a cabinet/

enclosure

8 | Autoclave | required on site | required in the building | required in the laboratory suite4 | double ended autoclave required in laboratory

System of work

9 | Access restricted to authorised personnel only | not required | required | required | required (via airlock key procedure)

10 | Biohazard sign on door | not required | required | required | required

11 | Specific measures to control aerosol dissemination | not required | required so as to minimise | required so as to prevent | required so as to prevent

12 | Shower | not required | not required | required where and to extent the risk assessment shows it is required | required

13 | Protective clothing | suitable protective clothing required | suitable protective clothing required | suitable protective clothing required; footwear required where and to extent the risk assessment shows it is required | complete change of clothing and footwear required before entry and exit

14 | Gloves | not required | required where and to extent the risk assessment shows they are required | required | required

15 | Efficient control of disease vectors (e.g. rodents and insects) which could disseminate GMMs | required where and to extent the risk assessment shows it is required | required | required | required

Waste

16 | Inactivation of GMMs in effluent from hand-washing sinks and showers and similar effluents | not required | not required | required where and to extent the risk assessment shows it is required | required

17 | Inactivation of GMMs in contaminated material and waste | required by validated means where and to extent the risk assessment shows it is required | required by validated means | required by validated means, with waste inactivated within the laboratory suite | required by validated means, with waste inactivated within the laboratory

Other measures

18 | Laboratory to contain its own equipment | not required | not required | required, so far as is reasonably practicable | required

19 | An observation window or alternative is to be present so that occupants can be seen | required where and to extent the risk assessment shows it is required | required where and to extent the risk assessment shows it is required | required where and to extent the risk assessment shows it is required | required

20 | Safe storage of GMMs | required where and to extent the risk assessment shows it is required | required | required | secure storage required

21 | Written records of staff training | not required | required where and to extent the risk assessment shows it is required | required | required

Table 1b

Containment measures applicable to contained use involving micro-organisms in plant growth facilities (to be read with Table 1a as indicated in paragraph 3(b) of Part 1)

1 A permanent structure refers to a fixed structure with walls, a roof and a floor. Where the permanent structure is a greenhouse, that structure must also have a continuous waterproof covering and self-closing lockable outer doors, and be located on a site designed to prevent the entry of surface run-off water.

Containment Measures | Containment Levels | Additional/ modification

1 | 2 | 3 | 4

Facilities

1 | Permanent structure1 | required where and to extent the risk assessment shows it is required | required | required | required | modification

Equipment

2 | Entry via a separate room with two interlocking doors | not required | required where and to extent the risk assessment shows it is required | required where and to extent the risk assessment shows it is required | required (via airlock key procedure) | additional

3 | Control of contaminated run-off water | required where and to extent the risk assessment shows it is required | required so as to minimise run-off | required so as to prevent run-off | required so as to prevent run-off | additional

System of work

4 | Effective control of disease vectors such as insects, rodents and arthropods which could disseminate GMMs | required | required | required | required | additional

5 | Effective control of pollen, seeds and other plant material which could disseminate GMMs | required where and to extent the risk assessment shows it is required | required so as to minimise dissemination | required so as to prevent dissemination | required so as to prevent dissemination | additional

6 | Procedures for transfer of living material between the plant growth facilities, protective structure and laboratory must control dissemination of GMMs | required so as to minimise dissemination | required so as to minimise dissemination | required so as to prevent dissemination | required so as to prevent dissemination | additional

Table 1c

Containment measures applicable to contained use involving micro-organisms in animal units (to be read with Table 1a as indicated in paragraph 3(c))

1 “animal unit” means a building, or separate area within a building, containing an animal facility and other areas including changing rooms, showers, autoclaves and food storage areas.

2 “animal facility” means a facility normally used to house stock, breeding or experimental animals or one which is used for the performance of minor surgical procedures on animals.

3 “isolators” means transparent boxes where small animals are contained within or outside a cage; for large animals, isolated rooms may be more appropriate.

Containment Measures | Containment Levels | Additional/ modification

1 | 2 | 3 | 4

Facilities

1 | Isolation of animal unit1 | required where and to extent the risk assessment shows it is required | required | required | required | modification

2 | Animal facilities2 separated by lockable doors | required where and to extent the risk assessment shows it is required | required | required | required | additional

3 | Animal facilities (cages, etc.) designed to facilitate decontamination (waterproof and easily washable material) | required where and to extent the risk assessment shows it is required | required where and to extent the risk assessment shows it is required | required | required | additional

4 | Floor, walls and ceiling easily washable | required where and to extent the risk assessment shows it is required | required for floor | required for floor and walls | required for floor, walls and ceiling | modification

5 | Appropriate filters on isolators or isolated rooms3 | not required | required where and to extent the risk assessment shows it is required | required | required | additional

6 | Appropriate barriers at the room exit, and at drains or ventilation duct work | required | required | required | required | additional

7 | Animals kept in appropriate containment facilities, such as cages, pens or tanks but not isolators | required where and to extent the risk assessment shows it is required | required where and to extent the risk assessment shows it is required | required where and to extent the risk assessment shows it is required | required where and to extent the risk assessment shows it is required | additional

8 | Animals kept in isolators | not required | required where and to extent the risk assessment shows it is required | required | required | modification

Table 2

Containment measures applicable to contained use involving micro-organisms in premises other than those referred to in Tables 1a, 1b and 1c

Containment Measure | Containment Levels

1 | 2 | 3 | 4

General

1 | Viable micro-organisms must be contained in a system which separates the process from the workplace and wider environment (closed system) | required where and to extent the risk assessment shows it is required | required | required | required

2 | Closed systems located within a controlled area | not required | required where and to extent the risk assessment shows it is required | required | required

3 | Control of exhaust gases from the closed system | not required | required so as to minimise release | required so as to prevent release | required so as to prevent release

4 | Control of aerosols during sample collection, addition of material to a closed system or transfer of material to another closed system | required where and to extent the risk assessment shows it is required | required so as to minimise release | required so as to prevent release | required so as to prevent release

5 | Inactivation of bulk culture fluids before removal from the closed system | required where and to extent the risk assessment shows it is required | required by validated means | required by validated means | required by validated means

6 | Seals must be designed so as to minimise or prevent release | not required | required so as to minimise release | required so as to prevent release | required so as to prevent release

7 | The controlled area designed to contain spillage of the entire contents of the closed system | required where and to extent the risk assessment shows it is required | required where and to extent the risk assessment shows it is required | required | required

8 | The controlled area sealable to permit fumigation | not required | required where and to extent the risk assessment shows it is required | required where and to extent the risk assessment shows it is required | required

9 | Biohazard signs posted | not required | required | required | required

Equipment

10 | Entry via airlock | not required | not required | required where and to extent the risk assessment shows it is required | required

11 | Surfaces resistant to water, acids, alkalis, solvents, disinfectants and decontamination agents and easy to clean | required for any bench | required for any bench | required for any bench and floor | required for any bench, floor, ceilings and walls

12 | Specific measures to ventilate adequately the controlled areas in order to minimise air contamination | required where and to extent the risk assessment shows they are required | required where and to extent the risk assessment shows they are required | required where and to extent the risk assessment shows they are required | required

13 | The controlled area maintained at an air pressure negative to the immediate surroundings | not required | not required | required where and to extent the risk assessment shows it is required | required

14 | Extract and input air from the controlled area must be HEPA filtered | not required | not required | required for extract air; required where and to extent the risk assessment shows it is required for input air | required for input and extract air

System of work

15 | Access restricted to authorised personnel only | not required | required | required | required

16 | Personnel must shower before leaving the controlled area | not required | not required | required where and to extent the risk assessment shows it is required | required

17 | Personnel must wear protective clothing | work clothing required | work clothing required | required | complete change required before exit and entry

18 | Written procedures and records of staff training | not required | required where and to extent the risk assessment shows they are required | required | required

Waste

19 | Inactivation of GMMs in effluent from hand-washing sinks and showers or similar effluents | not required | not required | required where and to extent the risk assessment shows it is required | required

20 | Inactivation of GMMs in contaminated material and waste including those in process effluent before final discharge | required by validated means where and to extent the risk assessment shows it to be required | required by validated means | required by validated means | required by validated means

Changed on 1 Oct 2014

1 “isolation” means, in relation to a laboratory, separation of the laboratory from other areas in the same building, or being in a separate building. 2 Entry must be through an airlock which is a chamber isolated from the laboratory. The clean side of the airlock must be separated from the restricted side by changing or showering facilities and preferably by interlocking doors. 3 Where viruses are not retained by the HEPA filters, extra requirements will be necessary for extract air. 4 Where the autoclave is outside the laboratory in which the contained use is being undertaken, but within the laboratory suite, there must be validated procedures for the safe transfer of material into that autoclave, which provide a level of protection equivalent to that which would be achieved by having an autoclave in that laboratory.

(1) “isolation” means, in relation to a laboratory, separation of the laboratory from other areas in the same building, or being in a separate building. (2) Entry must be through an airlock which is a chamber isolated from the laboratory. The clean side of the airlock must be separated from the restricted side by changing or showering facilities and preferably by interlocking doors. (3) Where viruses are not retained by the HEPA filters, extra requirements will be necessary for extract air. (4) Where the autoclave is outside the laboratory in which the contained use is being undertaken, but within the laboratory suite, there must be validated procedures for the safe transfer of material into that autoclave, which provide a level of protection equivalent to that which would be achieved by having an autoclave in that laboratory.

1 | Laboratory suite: isolationisolation1 | not required | not required | required | required

4 | Entry to laboratory via airlockairlock2 | not required | not required | required where and to extent the risk assessment shows it is required | required

6 | Extract and input air from the laboratory must be HEPA filtered | not required | not required | HEPA filters required for extract air except for activities where transmission does not occur by the airborne route | HEPA filters required for input and extract airair3

8 | Autoclave | required on site | required in the building | required in the laboratory suitesuite4 | double ended autoclave required in laboratory

1 A permanent structure refers to a fixed structure with walls, a roof and a floor. Where the permanent structure is a greenhouse, that structure must also have a continuous waterproof covering and self-closing lockable outer doors, and be located on a site designed to prevent the entry of surface run-off water. Containment Measures | Containment Levels | Additional/ modification

(1) A permanent structure refers to a fixed structure with walls, a roof and a floor. Where the permanent structure is a greenhouse, that structure must also have a continuous waterproof covering and self-closing lockable outer doors, and be located on a site designed to prevent the entry of surface run-off water.

1 | Permanent structurestructure1 | required where and to extent the risk assessment shows it is required | required | required | required | modification

1 “animal unit” means a building, or separate area within a building, containing an animal facility and other areas including changing rooms, showers, autoclaves and food storage areas. 2 “animal facility” means a facility normally used to house stock, breeding or experimental animals or one which is used for the performance of minor surgical procedures on animals. 3 “isolators” means transparent boxes where small animals are contained within or outside a cage; for large animals, isolated rooms may be more appropriate. Containment Measures | Containment Levels | Additional/ modification

(1) “animal unit” means a building, or separate area within a building, containing an animal facility and other areas including changing rooms, showers, autoclaves and food storage areas. (2) “animal facility” means a facility normally used to house stock, breeding or experimental animals or one which is used for the performance of minor surgical procedures on animals. (3) “isolators” means transparent boxes where small animals are contained within or outside a cage; for large animals, isolated rooms may be more appropriate.

1 | Isolation of animal unitunit1 | required where and to extent the risk assessment shows it is required | required | required | required | modification 2 | Animal facilitiesfacilities2 separated by lockable doors | required where and to extent the risk assessment shows it is required | required | required | required | additional

5 | Appropriate filters on isolators or isolated roomsrooms3 | not required | required where and to extent the risk assessment shows it is required | required | required | additional