, The Control of Substances Hazardous to Health Regulations 2002

In forcesince 21 Nov 2002

Containment measures | Containment levels

2 | 3 | 4

1 | The workplace is to be separated from any other activities in the same building. | No | Yes | Yes

2 | Input air and extract air to the workplace are to be filtered using HEPA or equivalent. | No | Yes, on extract air | Yes, on input and double on extract air

3 | Access is to be restricted to authorised persons only. | Yes | Yes | Yes, via air-lock key procedure

4 | The workplace is to be sealable to permit disinfection. | No | Yes | Yes

5 | Specified disinfection procedure. | Yes | Yes | Yes

6 | The workplace is to be maintained at an air pressure negative to atmosphere. | No | Yes | Yes

7 | Efficient vector control eg rodents and insects. | Yes, for animal containment | Yes, for animal containment | Yes

8 | Surfaces impervious to water and easy to clean. | Yes, for bench | Yes, for bench and floor (and walls for animal containment) | Yes, for bench, floor, walls and ceiling

9 | Surfaces resistant to acids, alkalis, solvents, disinfectants. | Yes, for bench | Yes, for bench and floor (and walls for animal containment) | Yes, for bench, floor, walls and ceiling

10 | Safe storage of biological agents. | Yes | Yes | Yes, secure storage

11 | An observation window, or alternative, is to be present, so that occupants can be seen. | No | Yes | Yes

12 | A laboratory is to contain its own equipment. | No | Yes, so far as is reasonably practicable | Yes

13 | Infected material, including any animal, is to be handled in a safety cabinet or isolator or other suitable containment. | Yes, where aerosol produced | Yes, where aerosol produced | Yes

14 | Incinerator for disposal of animal carcases. | Accessible | Accessible | Yes, on site

Changed on 21 Nov 2002

1 | Viable micro-organisms should be contained in a system which physically separates the process from the environment (closed system). | YesThe workplace is to be separated from any other activities in the same building. | No | Yes | Yes 2 | Exhaust gases from the closed system should be treated so as to— | Minimise release | Prevent release | Prevent release 3 | Sample collection, addition of materials to a closed system and transfer of viable micro-organisms to another closed system, should be performed so as to— | Minimise release | Prevent release | Prevent release 4 | Bulk culture fluids should not be removed from the closed system unless the viable micro-organisms have been— | Inactivated by validated means | Inactivated by validated chemical or physical means | Inactivated by validated chemical or physical means 5 | Seals should be designed so as to— | Minimise release | Prevent release | Prevent release 6 | Closed systems should be located within a controlled area— | Optional | Optional Input air and extract air to the workplace are to be filtered using HEPA or equivalent. | No | Yes, on extract air | Yes, on input and double on extract air 3 | Access is to be restricted to authorised persons only. | Yes | Yes | Yes, via air-lock key procedure 4 | The workplace is to be sealable to permit disinfection. | No | Yes | Yes 5 | Specified disinfection procedure. | Yes | Yes | Yes 6 | The workplace is to be maintained at an air pressure negative to atmosphere. | No | Yes | Yes 7 | Efficient vector control eg rodents and insects. | Yes, for animal containment | Yes, and purpose-built (a) biohazard signs should be posted; | Optional | Yes | Yes (b) access should be restricted to nominated personnel only; | Optional | Yesfor animal containment | Yes 8 | Surfaces impervious to water and easy to clean. | Yes, for bench | Yes, via air-lock (c) personnel should wear protective clothing; | Yes, work clothing for bench and floor (and walls for animal containment) | Yes | Yes, a complete change (d) decontamination and washing facilities should be provided for personnel; , for bench, floor, walls and ceiling 9 | Surfaces resistant to acids, alkalis, solvents, disinfectants. | Yes | Yes, for bench | Yes (e) personnel should shower before leaving the controlled area; | No | Optional | Yes (f) effluent from sinks and showers should be collected and inactivated before release; | No | Optional, for bench and floor (and walls for animal containment) | Yes, for bench, floor, walls and ceiling 10 | Safe storage of biological agents. | Yes | Yes | Yes (g) the controlled area should be adequately ventilated to minimise air contamination; | Optional | Optional | Yes (h) the controlled area should be maintained at an air pressure negative to atmosphere; , secure storage 11 | An observation window, or alternative, is to be present, so that occupants can be seen. | No | OptionalYes | Yes (i) input and extract air to the controlled area should be HEPA filtered; 12 | A laboratory is to contain its own equipment. | No | Optional Yes, so far as is reasonably practicable | Yes (j) the controlled area should be designed to contain spillage of the entire contents of closed system; | Optional | Yes13 | Infected material, including any animal, is to be handled in a safety cabinet or isolator or other suitable containment. | Yes (k) the controlled area should be sealable to permit fumigation. | No | Optional | Yes 7 | Effluent treatment before final discharge. | Inactivated by validated means | Inactivated by validated chemical or physical means | Inactivated by validated physical means, where aerosol produced | Yes, where aerosol produced | Yes 14 | Incinerator for disposal of animal carcases. | Accessible | Accessible | Yes, on site

  1. 21 Nov 2002 – todayin force
  2. 24 Oct 2002 – 21 Nov 2002earlier wording