The basic requirement is to obtain a representative and homogeneous laboratory sample.

The analyst shall ensure that samples do not become contaminated during sample preparation. Containers have to be rinsed with high purity acetone or hexane (p.A., HLPC grade or equivalent) before use to minimise the risk of contamination. Wherever possible, apparatus coming into contact with the sample shall be made of inert materials e.g. glass or polished stainless steel. Plastics such as polypropylene are to be avoided, because the analyte can adsorb onto these materials.

All of the sample material received by the laboratory is to be used for the preparation of test material. Only very finely homogenised samples give reproducible results.

There are many satisfactory specific sample preparation procedures which may be used.

For the purpose of this Annex the following definitions shall apply:

Solvent free massThe mass of the material after abstraction of the solvent, which normally is water.

Volatile fractionThe part of the solvent free mass, which is volatile and analysable by gas chromatography.

Identification of a primary productResults of a descriptive analysis, which identifies substances present in the primary product.

Characterisation of a primary productIdentification of the major physical-chemical fractions and quantification and identification of the chemical constituents.

LOQLimit of Quantification

LODLimit of Detection

SiThe single-laboratory standard deviation, calculated from results generated under repeatability conditions as defined in ISO standard 5725-1 [ISO 5725-1: Accuracy (trueness and precision) of measurement methods and results — Part 1: General principles and definitions. Geneva, 1994.] (= repeatability standard deviation estimated in a single laboratory approach according to the Harmonized Guidelines for Single-Laboratory Validation of Methods of Analysis [Thompson, M., S.L.R. Ellison, and R. Wood, Harmonized Guidelines for Single-Laboratory Validation of Methods of Analysis. Pure and Applied Chemistry, 2002. 74(5): pp. 835-855.]).

SrThe average within laboratory standard deviation, calculated from results generated under repeatability conditions as defined in ISO standard 5725-1 [ISO 5725-1: Accuracy (trueness and precision) of measurement methods and results — Part 1: General principles and definitions. Geneva, 1994.] in a collaborative trial with a minimum of eight laboratories conducted according to the Protocol for the Design, Conduct and Interpretation of Method-Performance Studies [Horwitz, W., Protocol for the design, conduct and interpretation of method-performance studies. Pure and Applied Chemistry, 1995. 67(2): pp. 331-343.].

SRThe between laboratory standard deviation, calculated from results under reproducibility conditions as defined in ISO standard 5725-1 [ISO 5725-1: Accuracy (trueness and precision) of measurement methods and results — Part 1: General principles and definitions. Geneva, 1994.] and according to the Protocol for the Design, Conduct and Interpretation of Method-Performance Studies [Horwitz, W., Protocol for the design, conduct and interpretation of method-performance studies. Pure and Applied Chemistry, 1995. 67(2): pp. 331-343.].

RSDiRelative single-laboratory repeatability standard deviation (Si expressed in percent of the measured value),

RSDrRelative average repeatability standard deviation (Sr expressed in percent of the measured value),

RSDRRelative reproducibility standard deviation (SR expressed in percent of the measured value).

Without prejudice to Article 11 of Regulation (EC) No 882/2004, the validated method for identification and characterisation to be selected by the laboratory shall comply with the quality criteria indicated in Tables 1 and 2.

Parameter | Value/Comment

Solvent free mass | At least 50 % by mass shall be identified and quantified

Volatile fraction | At least 80 % by mass shall be identified and quantified

[Over the whole analytical range.] [The RSDi, RSDr and RSDR values are relatively high due to the low stability of the analytes in primary smoke condensate.] [Corrected for recovery.]

Analyte(s) PAH | RSDi | RSDr | RSDR | LOD | LOQ | Analytical range | Recovery

% | % | % | μg/kg | μg/kg | μg/kg | %

benzo[a]pyrene | 20 | 20 | 40 | 1,5 | 5,0 | 5,0-15 | 75-110

benzo[a]anthracene | 20 | 20 | 40 | 3,0 | 10 | 10-30 | 75-110

cyclopenta[cd]pyrene

dibenzo[a,e]pyrene

dibenzo[a,i]pyrene

dibenzo[a,h]pyrene

35 | 35 | 70 | 5,0 | 15 | 15-45 | 50-110

chrysene

5-methylchrysene

benzo[b]fluoranthene

benzo[j]fluoranthene

benzo[k]fluoranthene

indeno[123-cd]pyrene

dibenzo[a,h]anthracene

benzo[ghi]perylene

dibenzo[a,l]pyrene

25 | 25 | 50 | 5,0 | 15 | 10-30 | 60-110